Product Description
Wiring configuration
1P2W, 1P3W, 3P3W, 3P4W
Measurements and parameters
Ior : Leakage current (Trms) with resistive components only.
Io : Leakage current (Trms) with basic wave of 40 – 70Hz.
Iom : Leakage current (Trms) including harmonic components
V : Reference voltage (Trms) with basic wave of 40 – 70Hz
Vm : Reference voltage (Trms) including harmonic components.
R : Insulation resistance, Frequency(Hz), Phase angle(θ)
Other functions
Digital output, Print screen, Back light, Data hold
Recording Interval
200/400ms/1/5/15/30s/1/5/15/30m/1/2hours
Ior
Range
10.000/100.00/1000.0mA/10.000A/AUTO
Accuracy
For reference voltages of sine wave 40 – 70Hz and 90V Trms or higher,
±0.2%rdg±0.2%f.s. + clamp sensor amplitude accuracy + error
of phase accuracy*
(phase error)
* add ±2.0%rdg to measured Io value when using Ior leakage
clamp sensor.
(θ: within the accuracy of reference voltage/ current phase
difference ±1.0°)
Allowable input
1% – 110% (Trms) of each range, and 200% (peak) of the range
Display range
0.15% – 130% (display “0” for less than 0.15%, “OL” if the range
is exceeded)
Io *Range, Allowable input and Display Range are the same as Ior .
Accuracy
±0.2%rdg±0.2%f.s.+ clamp sensor amplitude accuracy
Iom *Range, Allowable input and Display Range are the same as Ior .
Accuracy
±0.2%rdg±0.2%f.s.+ clamp sensor amplitude accuracy
Measurement method
0.9V – 1100.0V Trms (display “0” for less than 0.9V, “OL” if the
range is exceeded)
Phase angle(θ)
Display range
0.0° – ±180.0° (regarding the phase of reference voltage as 0.0°)
Accuracy
Within ±0.5° for the inputs of 10% or higher of leakage current
range, sine wave 40 – 70Hz, reference voltage of 90V Trms or
higher.
Frequency meter range
40 – 70Hz
External supply
AC100 – 240V(50/60Hz) 7VA max
Power source
LR6(AA)(1.5V) × 6 (Battery life approx. 11h)
Display / update period
160 × 160dots, FSTN monochrome display / 500ms
PC card interface
SD card (2GB) *standard accessory
PC communication interface
USB Ver2.0
Temperature and humidity range
23±5°C, less than 85%RH(without condensation)
Operating temperature and humidity range
-10 – 50°C less than 85%RH(without condensation)
Storage temperature and humidity range
-20 – 60°C less than 85%RH(without condensation)
Applicable Standards
IEC61010-1 CATⅣ 300V , CATⅢ 600V Pollution degree 2
IEC61010-2-030 , IEC61010-031 , IEC61326
Dimension/Weight
Dimension/Weight
Included accessories
7273(Voltage test lead)
8262(AC adapter)
7278(Earth cable)
7219(USB cable)
8326-02(SD card 2GB)
9125(Carrying case)
Instruction manual, Cable marker, Software installation manual
LR6(AA) × 6
KEW Windows for KEW 5050(software)
Optional accessories
8177(Ior Leakage clamp sensor 10A type Ø40mm)
8178(Ior Leakage clamp sensor 10A type Ø68mm)
8329(Power supply adapter)
Shows insulation resistance (R) values determined by the following formula.
V: Reference voltage/ Ior: Leakage current
Displayed value is just for reference since the measurement method differs from insulation resistance testers and may not be consistent with each other.