Back

KEW 5010

Price : 0/Piece(s)
Minimum Order : 1 Piece(s)
Production Capacity : 1 Piece(s)/week
Kyoritsu KEW India Instruments Pvt. Ltd.

Ahmedabad

Visit Exhibitor Profile
Product Description

KEW 5010

Photos

  • Features

  • 3 channel inputs for the simultaneous recording of Leakage Current and Load Current Large capacity for storing 60,000 data points
  • Lowpass Filter will filter out the harmonics
  • LED flickers when the preset current value is exceeded
  • CALL:Confirmation of recorded data
  • Selection of One-time mode or Endless mode
  • Non Volatile Memory
  • Battery power indicator
  • The userfriendly PC software “KEW LOG Soft 2” is supplied
  • Continuous measuring time:Approx. 10 days (Alkaline Battery)
  • Product Description

  • 3 channel inputs for the simultaneous recording of Leakage Current, Load Current and Voltage
  • Large Capacity for storing 60,000 data points :- 60,000 data points can be recorded when 1ch is used, and when all the three channels are used, 20,000 data points per channel can be recorded
  • Lowpass Filter will filter out the harmonics (Cutoff Frequency=Approx. 160Hz)
  • LED flickers when the preset current value is exceeded
  • CALL: Confirmation of recorded data :- The following can be displayed: number of recorded data points, (max+ min+ peak) value for each channel complete with time/date information in the Normal recording mode. (Detected values (i.e. when values are outside preset limits ) can be displayed in other recording modes). RECALL: The last 10 recorded data points including time/date can be recalled on the logger display
  • Selection of One-time mode or Endless mode :- One-time on: Recording will stop when memory is used up. One-time off: Overwrite the old data, and store the latest data
  • Non Volatile Memory :- Recorded data will be retained even if the batteries are exhausted or replaced due to the presence of a nonvolatile memory (guaranteed for 10 years)
  • Battery power indicator :- ndicates battery voltage in 4-levels (It is possible to use the logger for a further approx. 24 hours even after the warming symbol is flashing)
  • The userfriendly PC software “KEW LOG Soft 2” is supplied :- Supplied with the user friendly software “KEW LOG Soft 2”. This permits editing, analysis and graphical display of data. The recorded data is downloadable onto a PC via USB cable.
  • Continuous measuring time: Approx. 10 days (Alkaline Battery)
  • Normal Recording Mode (AC 50/60Hz, Sine wave, Input: 10% or more of the range at CH1)

    Range
    RMS Accuracy
    100.0mA
    ±2.0%rdg±0.9%f.s.+ Accuracy of Senso
    Other ranges
    ±1.5%rdg±0.7%f.s.+ Accuracy of Sensor
    Crest Factor
    2.5 or less: RMS accuracy (sine) + 2%rdg + 1%f.s.
    Max, Min and Instant Peak values in Normal Recording mode are just reference values: their accuracies aren’t guaranteed.

    Trigger Recording Mode (AC 50/60Hz sine wave)

    Range
    Accuracy
    100.0mA
    ±3.5%rdg±2.2%f.s.+ Accuracy of Sensor
    Other ranges
    ±3.0%rdg±2.0%f.s.+ Accuracy of Sensor

    Capture Recording Mode

    Range
    Accuracy
    100.0mA
    100.0mA
    Other ranges
    ±2.5%rdg±1.5%f.s.+ Accuracy of Sensor
    Recording mode
    Normal, Trigger, Capture
    Operating system
    Successive Approximation (CH1 single synchronized sampling)
    Rated max. working voltage
    AC 9.9Vrms, 14V peak value
    Number of input channel
    3ch
    Measuring method
    True RMS
    RMS measuring interval
    approx. 100ms
    Sampling interval
    Normal/Trigger mode : approx. 1.65ms/CH Capture mode : approx. 0.55ms (waveform: at every 1.1ms)
    Low battery warning
    Battery mark display (in 4 levels)
    Over-range indication
    “OL” mark is displayed when exceeding the measuring range
    Auto power off
    Power-off function operates automatically after a switch remains for 3min. (when recording is stopped)
    Possible measurement time
    Approx. 10 days (with alkaline LR6 batteries)
    Applicable standards
    IEC 61010-1:2001 CAT III 300V Pollution degree2 IEC 61326 (EMC standard)
    Dimensions
    111(L)×60(W)×42(D)mm
    Weight
    Approx. 265g
    Included Accessories
    Alkaline battery LR6×4pcs, 9118 (Carrying case) PC software “KEW LOG Soft 2”, 7148 (USB cable) Instruction manual, Quick manual, Install manual, USB Notice sheet
    Optional Accessories
    8146/8147/8148 (Leakage & Load current Clamp Sensor), 8121/8122/8123 (Load current Clamp Sensor), 8129/8130 (Flexible Clamp Sensor), 8320 (AC Adapter), 9135 (Carrying bag), 9119 (Hard carry case), 7185 (Extension cable for sensor)