Product Description
Earth resistance Auto range
20.00/200.0/1500Ω
±1.5%±0.05Ω(0.00 – 20.99Ω)*
±2%±0.5Ω(16.0 – 99.9Ω)
±3%±2Ω(100.0 – 209.9Ω)
±5%±5Ω(160 – 399Ω)
±10%±10Ω(400 – 599Ω)
AC current (50Hz/60Hz) Auto range
Values are displayed, but accuracy isn’t guaranted(600 – 1580Ω)
100.0/1000mA/10.00/30.0A
±2%±0.7mA(0.0 – 104.9mA)
±2%(80mA – 31.5A)
Operating indication
Earth resistance function :
Constant voltage injection
Current detection (Frequency : Approx.2400Hz)
Dual Integration
AC current function : Successive approximation
Over-range indication
“OL”is displayed when input exceeds the upper limit of a measuring range
Response time
Approx. 7 seconds (Earth resistance)
Approx. 2 seconds (AC current)
Sample rate
Approx. 1 times per second
Current consumption
Approx. 50mA (max.100mA)
Measurement time
Approx.12 hours (when R6 is used)
Approx.24 hours (when LR6 is used)
Auto power-off
Turns power off about 10 minutes after the last button operation.
Conductor size
Approx. Ø32mm
Applicable standards
IEC 61010-1 CAT Ⅳ 300V Pollution degree2
IEC 61010-2-032, IEC 61326 (EMC)
Power source
DC6V : R6(sizeAA manganese battery) x 4
or LR6 (sizeAA alkaline battery) x 4
Dimensions
246(L)×120(W)×54(D)mm
Weight
Approx. 780g (including batteries)
Included Accessories
R6 x 4, Instruction manual
8304 (Resister for operation check)
9166 (Carrying case[Hard])